Error/Event
Code
Probable Cause
Remedy
2
Event. Parameters have been set to factory defaults.
You can change the default settings in configuration mode see Communicating with Terminal Software (page Event. Wind calibration data has been lost.
WMT700 needs to be calibrated.
Contact Vaisala technical support.
10
Error. You have given an invalid value when using the
S command.
Check the allowed parameter values see E Configurable
Parameters (page Error. Parameter used with the
S command is unknown.
Check that you are using the correct parameter names, see
E.1
Configurable Parameters (page Error. Unknown command.
Check the allowed commands in A. Command Set for WMT700
(page Error. String before the end of command character is too long.
Check the command and parameter names that you are using.
You can also troubleshoot WMT700 as follows:
• Fetch the error and event messages with the
ERRORS command.
• Include diagnostics-related items in the data message.


Document code
Testsetup according to
More information
Optiontemperature range
Optionheating
Optionhardware interface
Optionunit used
Optionoutput configuration
Selectionoutput configuration
Optionadapter type
Optionaccessories
Order codedescription
Aout1 selected option
Output signal
Swnddiroffset,(angle)
Parameterdefinition
Power supplywire colors
Used heating option
Heating option
Wire type/cable length
Descriptionspare part item
Wmt700connector pin
Open commandto activate the command, press enter
Commanddescription
Wind_getfetches wind calibration information.more information
S command. power must be on for at least 5 seconds after the s
Startcommand.more information
Itemdescription
Bitdescription
G command to fetch the current wmt700 parameters. disable error messages with the command s messages,0
Errors. if all values are valid, wmt700 sends the following response:0,0,0,0you can also use the g
Data message number
Ws425 and sdi messages
Parametername
Parameterdefault
Start command.to stop measurement, use the stop
Wb,wfirmware
Wf output:wff6785672109876.5432.1k99xx5.21.2 ws425 f/g asos data messagewmt700 responds to the wa
Bytedescription
Parameterdefault value
Id character
Wx command with a character fixed-length data message.character
Start command. to stop measurement, use thestop
Problemprobable cause
Parameter name
Propertydescription/value
Analog outputs for wind direction
Itemorder code
Configuration mode / measurement
Settingparameter
Start command. you can stop the measurement with the stop
Example of the command and response:
Baud........................................................................ 105clearerr
S............................................................................... 103sleep
Gget message. get parameter.........................................................104h
Jjunction box...............................................................28l



Share with your friends:
1   ...   94   95   96   97   98   99   100   101   ...   125


The database is protected by copyright ©userg.info 2017
send message

    Main page

bosch
camera
chevrolet
epson
fiat
Harley Davidson
Honda
iphone
mitsubishi
nissan
Panasonic
Sony
volvo
xiaomi
Xperia
yamaha