Problem
Probable Cause
Remedy
Data messages are not in the expected format.
The selected data message is not correct.
If you are using automatic messages check the selected data message with the
G command. If necessary, set anew value for the autoSend parameter. If you are using polling Check that you are using the correct data message number in the polling command,
see Table 44 (page Some items are missing from the data messages.
The configured data message does not contain all the required items.
Define the data message again with the required items. See Parameter Handling Commands
(page Configuration commands do not work.
WMT700 is in measurement mode.
Switch from measurement mode to configuration mode. See Communicating with Terminal
Software (page 99)
WMT700 sends an error message as a response to a command.
For probable causes, see
7.2 Error and
Event Messages (page For remedies, see
7.2 Error and
Event Messages (page No signal is present on analog outputs.
Analog output has been disabled.
Enable analog output with the aout1_mode and aout2_mode parameters. See D. Configuration
Parameter Descriptions
(page Automatic data messages are not received although parameters have been set correctly.
WMT700 is not in continuous measurement mode.
Initiate continuous measurement with
START command. See START — Start Continuous
Measurement (page Data from WMT700 is temporarily lost.
The data logger and wind sensor are not in the same operating mode (polling or automatic transmission mode).
Make sure that the data logger and wind sensor are both in either polling or automatic transmission mode.
Data messages are not received.
Baud rates of the data logger and wind sensor do not match.
Change the baud rate so that it is the same in the data logger and wind sensor.
Parameter does not change after S command has been given.
The parameter requires reset before it becomes valid.
Check D. Configuration Parameter
Descriptions (page 187)
if reset is required before the parameter is changed.
Inconsistent serial communication or missing data.
Pin contacts of the connector have become oxidized or the connector is not firmly attached.
Change the WMT700 cable.
Retrofit installation of
WMT700 is not possible.
You do not have the correct mounting kit for the retrofit installation.
Check that you are using the correct mounting adapter with your mounting kit. See F. WMT700
Accessories (page 197)
WMT700 User Guide
M211095EN-H
168




Document code
Testsetup according to
More information
Optiontemperature range
Optionheating
Optionhardware interface
Optionunit used
Optionoutput configuration
Selectionoutput configuration
Optionadapter type
Optionaccessories
Order codedescription
Aout1 selected option
Output signal
Swnddiroffset,(angle)
Parameterdefinition
Power supplywire colors
Used heating option
Heating option
Wire type/cable length
Descriptionspare part item
Wmt700connector pin
Open commandto activate the command, press enter
Commanddescription
Wind_getfetches wind calibration information.more information
S command. power must be on for at least 5 seconds after the s
Startcommand.more information
Itemdescription
Bitdescription
G command to fetch the current wmt700 parameters. disable error messages with the command s messages,0
Errors. if all values are valid, wmt700 sends the following response:0,0,0,0you can also use the g
Data message number
Ws425 and sdi messages
Parametername
Parameterdefault
Start command.to stop measurement, use the stop
Wb,wfirmware
Wf output:wff6785672109876.5432.1k99xx5.21.2 ws425 f/g asos data messagewmt700 responds to the wa
Bytedescription
Parameterdefault value
Id character
Wx command with a character fixed-length data message.character
Start command. to stop measurement, use thestop
Problemprobable cause
Error/eventcode
Parameter name
Propertydescription/value
Analog outputs for wind direction
Itemorder code
Configuration mode / measurement
Settingparameter
Start command. you can stop the measurement with the stop
Example of the command and response:
Baud........................................................................ 105clearerr
S............................................................................... 103sleep
Gget message. get parameter.........................................................104h
Jjunction box...............................................................28l



Share with your friends:
1   ...   92   93   94   95   96   97   98   99   ...   125


The database is protected by copyright ©userg.info 2017
send message

    Main page

bosch
camera
chevrolet
epson
fiat
Harley Davidson
Honda
iphone
mitsubishi
nissan
Panasonic
Sony
volvo
xiaomi
Xperia
yamaha