User Guide M211095en-h



Download 8.38 Kb.
View original pdf
Page69/125
Date24.02.2019
Size8.38 Kb.
1   ...   65   66   67   68   69   70   71   72   ...   125
Parameter
Name
Default Value
Allowed Values
Description of How to Emulate WS425
aout1err
1000 0 …. Set 1 for voltage output and 625 for frequency output for similar operation as with WS425.
aout1_g
1 0 …. Gain for AOUT1. Set 0.017895 for compatibility with WS425 voltage output. Set 11.18 for compatibility with WS425 frequency output.
aout2_g
1 0 …. Gain for AOUT2. Set 0.0027855 for compatibility with WS425 potentiometer output.
aout1mode
3 0 = Current = Voltage = Frequency = Disabled
Analog output mode for AOUT1. Set 1 for WS425
voltage output compatibility. 2 for WS425
frequency output compatibility.
aout2mode
7 4 = Current = Voltage = Potentiometer
7 = Disabled
Analog output mode for AOUT2.Set 6 WS425
potentiometer output compatibility.
aout1_oaou to Offset for AOUT1 and AOUT2.Set output offset to
0.


Document code
Testsetup according to
More information
Optiontemperature range
Optionheating
Optionhardware interface
Optionunit used
Optionoutput configuration
Selectionoutput configuration
Optionadapter type
Optionaccessories
Order codedescription
Aout1 selected option
Output signal
Swnddiroffset,(angle)
Parameterdefinition
Power supplywire colors
Used heating option
Heating option
Wire type/cable length
Descriptionspare part item
Wmt700connector pin
Open commandto activate the command, press enter
Commanddescription
Wind_getfetches wind calibration information.more information
S command. power must be on for at least 5 seconds after the s
Startcommand.more information
Itemdescription
Bitdescription
G command to fetch the current wmt700 parameters. disable error messages with the command s messages,0
Errors. if all values are valid, wmt700 sends the following response:0,0,0,0you can also use the g
Data message number
Ws425 and sdi messages
Parameterdefault
Start command.to stop measurement, use the stop
Wb,wfirmware
Wf output:wff6785672109876.5432.1k99xx5.21.2 ws425 f/g asos data messagewmt700 responds to the wa
Bytedescription
Parameterdefault value
Id character
Wx command with a character fixed-length data message.character
Start command. to stop measurement, use thestop
Problemprobable cause
Error/eventcode
Parameter name
Propertydescription/value
Analog outputs for wind direction
Itemorder code
Configuration mode / measurement
Settingparameter
Start command. you can stop the measurement with the stop
Example of the command and response:
Baud........................................................................ 105clearerr
S............................................................................... 103sleep
Gget message. get parameter.........................................................104h
Jjunction box...............................................................28l