The IQ Force must be located in a nonhazardous location during
the charging cycle. The IQ Force’s battery charger is not approved
for use in hazardous locations.

IQ Force™ Gas Detector Reference Manual
19
Alarms
2.5.6 Sensor over-range alarms
For all gases and vapors other than oxygen, the IQ Force will go into alarm and display “OL” (over limit) if a sensor is exposed to a concentration of gas that exceeds its established range. In the case of an LEL reading that exceeds 100% LEL, the LEL channel will be automatically disabled by the instrument and the alarm will latch (remain on) until the instrument is turned off. The IQ Force must be turned off, brought to an area that is known to be safe (containing 20.9% oxygen, 0% LEL, and 0 ppm toxic gases, and then turned back on. The display will show “---” and “OL” alternately in place of the sensor reading for any channel that has gone into over-range alarm.
Toxic channel “OL” alarms occur when HS readings go over 200 ppm or CO readings go over 1000 ppm. When the toxic channel records below the
“OL” levels, the ppm readings return to the display. Any channel that reads and “OL” level will require recalibration.
WARNING


Iq force™ gas detector
Contacting honeywell analytics
Technical services
Appendix a toxic gas measurement – warning, danger, stel,
Appendix d toxic sensor cross-sensitivity .............................. 52
Operating temperature
Multi-sensor capability
Table 2. alarm logic
Iq force portable gas detectors have been designed for the
In the event of an lel over-range alarm, the iq force must be
A sensor that cannot be calibrated or is found to be out of toler-
Do not use the pump for prolonged periods in an atmosphere
Case: the instrument is enclosed in a solid pc (polycarbonate) case with tpe (rubber) overmold.2. front face:
On/off mode button:
Battery charger connector:
Turning the iq force on
Notethe bump test interval and bump test due reminder are
Notesample probe assembly and tubing must be attached when
Always verify that all sensors present in the instrument are
Turning the iq force off
The iq force is designed to detect potentially life threatening
A combustible sensor over-range alarm indicates a potentially
Pc connection via infrared port
Error messages
Manual sample-draw kit
Motorized sample-draw pump
The canadian standards association (csa) requires the lel
Calibration/accuracy verification
Figure 1. bump test/span calibration setup
Fresh air/zero calibration
Possible causes and solutions
Fresh air/zero calibrations may only be performed in an
Span calibration
Noteonce the calibration cycle is completed, the iq force will
Possible causes of span calibration failure and remedies:
A sensor that cannot be calibrated or is found to be out of
Noteto enable the h
Notehoneywell analytics recommends regular verification of
Notethe bump test interval and bump test reminder are used
Toxic gas measurement – warning, danger, stel, and twa alarms
A.2 time weighted average (twa)
Appendix b calibration frequency rec-
Verification of accuracy & related terminology
Lengthening the intervals between calibration checks or bump tests
One button automatic full calibration and calibration docks
Iq force sensor information



Share with your friends:
1   ...   37   38   39   40   41   42   43   44   ...   87


The database is protected by copyright ©userg.info 2017
send message

    Main page

bosch
camera
chevrolet
epson
fiat
Harley Davidson
Honda
iphone
mitsubishi
nissan
Panasonic
Sony
volvo
xiaomi
Xperia
yamaha